FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 9 DEFIBRILLATOR/MONITOR
MDR report key: 1021593
·
Received March 25, 2008
Report
- Report Number
- 3015876-2008-00217
- Event Type
- Malfunction
- Date Received
- March 25, 2008
- Date of Event
- February 25, 2008
- Report Date
- February 25, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K881153
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED PROBLEM. THE BATTERY PACK WAS REPLACED AND PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS SENT BACK TO LOANER STATUS. PHYSIO FURTHER EVALUATED THE REPLACED BATTERY PACK AND DETERMINED THE ROOT CAUSE FOR THE REPORTED FAILURE WAS A DEAD CELL.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON BATTERY POWER. THIS IS A PHYSIO-CONTROL LOANER UNIT. THERE WAS NO PT INVOLVEMENT REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 9 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 9 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |