FDA Adverse Event Malfunction Summary report: N

LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1021575 · Received March 27, 2008

Report

Report Number
3015876-2008-00229
Event Type
Malfunction
Date Received
March 27, 2008
Date of Event
February 26, 2008
Report Date
February 29, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO REPLACED THE POWER SUPPLY ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WOULD NOT OPERATE ON BATTERY POWER. THERE WAS NO PT INVOLVEMENT REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA