MULTI-LINK RX VISION CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2008-00217
- Event Type
- Malfunction
- Date Received
- March 26, 2008
- Date of Event
- February 27, 2008
- Report Date
- February 29, 2008
- Manufacturer
- ABBOTT VASCULAR-CARDIAC THERAPIES
- Product Code
- MAF
- PMA / PMN Number
- PO20047
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
RESULTS AND CONCLUSIONS SUMMATION - PRODUCT PERFORMANCE ENGINEERING REVIEWED THE INCIDENT INFORMATION. FACTORS THAT CAN CONTRIBUTE TO STENT DISLODGEMENT DURING PREPARATION INCLUDED, BUT ARE NOT LIMITED TO, IMPROPER OR INADEQUATE CRIMPING AT THE TIME OF MANUFACTURE, POSITIVE PRESSURE DURING PREPARATION, NEGATIVE PRESSURE DURING SHEATH REMOVAL OR FORCED SHEATH REMOVAL, MISHANDLING, OR INTERACTIONS WITH OTHER DEVICES. HOWEVER, WITHOUT A RETURNED DEVICE FOR ANALYSIS, A DEFINITE ROOT CAUSE OF THE STENT DISLODGEMENT CANNOT BE DETERMINED.
REPORTING RATIONALE: MALFUNCTION REPORTING STATUS: STENT DISLODGEMENT HAS CAUSED OR CONTRIBUTED TO PATIENT INJURY PREVIOUSLY DEVICE ISSUE: STENT DISLODGEMENT IT WAS REPORTED THAT DURING PREPARATION, THE STENT DISLODGED ONTO THE TABLE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL PATIENT OR EVENT DETAILS ARE AVAILABLE.