FDA Adverse Event Malfunction Summary report: N

AFFINITY 3

MDR report key: 1021505 · Received March 26, 2008

Report

Report Number
1824206-2008-00044
Event Type
Malfunction
Date Received
March 26, 2008
Date of Event
August 30, 2007
Report Date
August 30, 2007
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. THE HILL-ROM TECHNICIAN ADJUSTED THE BRAKE/STEER CASTER TO REPAIR THE BED. THE AFFINITY SERVICE MANUAL (MAN 272) DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC. AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY. THE STEERING PEDAL SHOULD ALSO BE CHECKED FOR PROPER LOCKING WHEN ACTIVATED.

Description of Event or Problem · 1

ACCOUNT ALLEGED THE BRAKES WOULD NOT STAY ENGAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM RITTER 3700

Patients

Seq Age Sex Outcome Treatment
1 UNK