FDA Adverse Event
Malfunction
Summary report: N
AFFINITY 3
MDR report key: 1021499
·
Received March 26, 2008
Report
- Report Number
- 1824206-2008-00065
- Event Type
- Malfunction
- Date Received
- March 26, 2008
- Date of Event
- December 20, 2007
- Report Date
- December 20, 2007
- Manufacturer
- HILL-ROM RITTER
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. THE HILL-ROM TECHNICIAN ADJUSTED THE BRAKE/STEER CASTER TO REPAIR THE BED. THE AFFINITY SERVICE MANUAL (MAN272) DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC. AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY. THE STEERING PEDAL SHOULD ALSO BE CHECKED FOR PROPER LOCKING WHEN ACTIVATED.
Description of Event or Problem · 1
ALLEGED NO BRAKE ON THE BED WHEN THE BRAKE WAS APPLIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY 3 | BIRTHING BED | HDD | HILL-ROM RITTER | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |