FDA Adverse Event Malfunction Summary report: N

AFFINITY 3

MDR report key: 1021498 · Received March 26, 2008

Report

Report Number
1824206-2008-00064
Event Type
Malfunction
Date Received
March 26, 2008
Date of Event
November 27, 2007
Report Date
November 27, 2007
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD TECHNICIAN INDICATED THE BRAKES WERE FAULTY. THE HILL-ROM TECHNICIAN FOUND THE BRAKE CASTER WAS WORN AND THE CAM PIVOT BROKEN. THE PARTS WERE REPLACED TO REPAIR THE BED. THE AFFINITY SERVICE MANUAL DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC., AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY. THE STEERING PEDAL SHOULD ALSO BE CHECKED FOR PROPER LOCKING WHEN ACTIVATED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE BRAKES WERE FAULTY ON THIS BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY 3 BIRTHING BED HDD HILL-ROM RITTER 3700

Patients

Seq Age Sex Outcome Treatment
1 UNK