INRATIO
Report
- Report Number
- 2954730-2008-00148
- Event Type
- Malfunction
- Date Received
- March 26, 2008
- Date of Event
- February 29, 2008
- Report Date
- March 25, 2008
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE - 2008, INRATIO - 2.9, LAB - 2.0, MEAN - 2.5, CONFIDENCE LIMITS - 1.6 - 3.4. DATE - 2008, INRATIO - 3.9, LAB - 2.5, MEAN - 3.2, CONFIDENCE LIMITS - 1.9 - 4.6. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR BOTH DATA SETS, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING IS NOT REQUIRED AT THIS TIME.
CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE - 2008, INRATIO - 2.9, LAB - 2.0. DATE - 2008, INRATIO - 3.9, LAB - 2.5.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | KIT, SELF-TEST, ENGLISH | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |