FREESTYLE LIBRELINK
Report
- Report Number
- 2954323-2020-04438
- Event Type
- Injury
- Date Received
- June 30, 2020
- Date of Event
- June 22, 2020
- Report Date
- August 11, 2020
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- PZE
- PMA / PMN Number
- P160030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
ADC PRODUCT QUALITY ENGINEERING (PQE) INVESTIGATED THIS COMPLAINT IN WHICH THE USER REPORTED NOT BEING ABLE TO SCAN A LIBRE SENSOR WHEN USING A SAMSUNG S9 DEVICE WITH ANDROID 10 WHEN THE BATTERY WAS BELOW 80%. PQE ATTEMPTED TO REPLICATE THE COMPLAINT USING A SIMILAR CONFIGURATION AND WAS NOT ABLE TO REPLICATE THE ISSUE. RESEARCH & DEVELOPMENT ALSO TRIED TO REPLICATE THIS ISSUE USING A SAMSUNG 9 WITH DIFFERENT OS VERSIONS (8, 9, 10), A LIBRE SENSOR, AND DIFFERENT BATTERY LEVELS. THE SENSOR SCANNED SUCCESSFULLY WITHOUT ANY ISSUES UNDER ALL SCENARIOS. AS PQE WAS NOT ABLE TO REPLICATE THE USER COMPLAINT, THE COMPLAINT IS NOT CONFIRMED. NO MALFUNCTION OR PRODUCT DEFICIENCY WAS IDENTIFIED.
CUSTOMER REPORTED BEING UNABLE TO SCAN THE ADC FREESTYLE LIBRE SENSOR USING THE ANDROID LIBRE LINK APP 10 WHEN THE SENSOR BATTERY IF BELOW 80%. CUSTOMER FURTHER REPORTED BEING UNWELL AND EXPERIENCED HEADACHE AND WAS TREATED WITH SUGAR BY A NON-HCP. NO FURTHER TREATMENT WAS REPORTED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. THE DEVICE MFG DATE IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
CUSTOMER REPORTED BEING UNABLE TO SCAN THE ADC FREESTYLE LIBRE SENSOR USING THE ANDROID LIBRE LINK APP 10 WHEN THE SENSOR BATTERY IF BELOW 80%. CUSTOMER FURTHER REPORTED BEING UNWELL AND EXPERIENCED HEADACHE AND WAS TREATED WITH SUGAR BY A NON-HCP. NO FURTHER TREATMENT WAS REPORTED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 675746 | FREESTYLE LIBRELINK | DATA MANAGEMENT SYSTEM | PZE | ABBOTT DIABETES CARE INC | 71733-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Required Intervention |