Report

Report Number
2953200-2008-00171
Event Type
Injury
Date Received
March 28, 2008
Date of Event
February 25, 2008
Report Date
February 25, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
NIQ
PMA / PMN Number
P060033
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: RESULTS - FAILURE TO FOLLOW INSTRUCTIONS (THE INNER POUCH CONTAINS A WARNING LABEL (("THIS PROTECTIVE OUTER POUCH CONTAINS AN INNER POUCH WITH A NON-STERILE EXTERIOR")) IFU STATES THAT THE OUTER SURFACE OF THE INNER POUCH HAS A NON-STERILE EXTERIOR WITH CONTENTS THAT ARE STERILE). CONCLUSIONS - USER ERROR CONTRIBUTED TO EVENT (THE INNER POUCH CONTAINS A WARNING LABEL (("THIS PROTECTIVE OUTER POUCH CONTAINS AN INNER POUCH WITH A NON-STERILE EXTERIOR")) IFU STATES THAT THE OUTER SURFACE OF THE INNER POUCH HAS A NON-STERILE EXTERIOR WITH CONTENTS THAT ARE STERILE).

Description of Event or Problem · 1

A 2.5MM DIAMETER X 18MM LENGTH ENDEAVOR MX2 DRUG-ELUTING CORONARY STENT WAS SUCCESSFULLY IMPLANTED INTO A PT FOR THE TREATMENT OF AN UNK LESION. THE LESION MORPHOLOGY IS UNK. IT WAS REPORTED THAT THE INNER POUCH THAT IS IDENTIFIED AS HAVING A NON-STERILE EXTERIOR WITH CONTENTS THAT ARE STERILE; WAS PLACED IN THE STERILE FIELD AND WAS OPENED BY THE SCRUB TECH. THE ENDEAVOR DRUG-ELUTING STENT WAS INSERTED INTO THE PT AND WAS SUCCESSFULLY IMPLANTED AT THE LESION SITE. IT WAS REPORTED AFTER THE CASE, THAT THE NON-STERILE POUCH WAS PUT IN THE STERILE FIELD. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS DOING WELL AND WILL BE MONITORED BY THE PHYSICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDEAVOR ZOTAROLIMUS-ELUTING CORONARY STENT MX2 NIQ MEDTRONIC CARDIOVASCULAR GALWAY NA 0000603196

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention