FDA Adverse Event Injury Summary report: N

PRECISION

MDR report key: 1021223 · Received March 28, 2008

Report

Report Number
2029203-2008-00156
Event Type
Injury
Date Received
March 28, 2008
Date of Event
February 27, 2008
Report Date
February 27, 2008
Manufacturer
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
Product Code
LGW
PMA / PMN Number
P030017
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CHARGER WILL NOT BE RETURNED TO BOSTON SCIENTIFIC. THE PATIENT'S IMPROPER CHARGING TECHNIQUE CAUSED THE REPORTED EVENT. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

THE PHYSICIAN DISCOVERED REDNESS AND A SCAB AT THE IMPLANT SITE. THE PHYSICIAN BELIEVES THAT THE PATIENT RECEIVED A PRESSURE WOUND FROM THE CHARGER. THE PATIENT ADMITTED TO SLEEPING WHILE CHARGING. THE PRODUCT LABELING INSTRUCTS THE PATIENT NOT TO CHARGE WHILE SLEEPING. THE PHYSICIAN REVISED THE POCKET SITE AS A PRECAUTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION SPINAL CORD STIMULATOR CHARGER LGW BOSTON SCIENTIFIC NEUROMODULATION CORPORATION SC-5300 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention