FDA Adverse Event Injury Summary report: N

TRUFILL DCS ORBIT DETACHABLE COILS COMPLEX

MDR report key: 1021179 · Received March 28, 2008

Report

Report Number
1058196-2008-00088
Event Type
Injury
Date Received
March 28, 2008
Date of Event
October 22, 2007
Report Date
February 29, 2008
Manufacturer
CORDIS NEUROVASCULAR, INC.
Product Code
HCG
PMA / PMN Number
K053197
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RETURNED FOR EVALUATION AND TESTING; HOWEVER, THE ENGINEERING EVALUATION IS NOT COMPLETED. ADD'L INFO WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT.

Description of Event or Problem · 1

THE COIL STRETCHED/SEPARATED IN PT. THIS PT WAS UNDERGOING COIL EMBOLIZATION WITHIN THE BASILAR TIP ANEURYSM MEASURING 14MM X 12MM X 11MM. NO CALCIFICATION AND MINIMAL TORTUOSITY WAS PRESENT WITHIN THE VESSEL. SEVEN COILS WERE PLACED USING THE PROWLER SELECT PLUS 606-S255X MICROCATHETER WITHOUT DIFFICULTY. NO RESISTANCE FELT WHILE ADVANCING THE 8TH COIL INTO THE MICROCATHETER. THE COIL WAS BEING WITHDRAWN FOR REPOSITIONING IN THE ANEURYSM WHEN STRETCHED AND WAS HALF IN THE ANEURYSM AND HALF IN THE MICROCATHETER. THERE WAS ONE TO ONE RELATIONSHIP BETWEEN THE COIL AND DELIVERY TUBE VERIFIED WITH FLUORO PRIOR TO REPOSITIONING. CONTINUOUS FLUSH WAS MAINTAINED WITHIN THE MICROCATHETER. THE COIL WAS STRETCHED ALL THE WAY TO THE GROIN. AT THIS TIME, IT WAS FELT THAT THE COIL WOULD INCORPORATE ITSELF INTO THE WALL OF THE VESSEL, CAUSING THE PT NO HARM. THE PT WILL REMAIN ON PLAVIX FOR THREE MONTHS..ASA FOR LIFE. THERE WAS ADVERSE EFFECT TO THE PT. THERE WAS NO RESISTANCE BETWEEN THE GW AND THE MC WHEN ACCESSING THE TARGET SITE. PRODUCT WILL BE RETURNED FOR ANALYSIS. THE SECTION OF STRETCHED COIL THEY WERE ABLE TO REMOVE. THE DISTAL PART OF THE COIL IS IN THE ANEURYSM AND STRETCHED DOWN THE BASILAR ALL THE WAY TO THE GROIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRUFILL DCS ORBIT DETACHABLE COILS COMPLEX CNV DCS ORBIT (HCG) HCG CORDIS NEUROVASCULAR, INC. NA 13320306

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention COILS & MC