FDA Adverse Event Malfunction Summary report: N

FREESTYLE FLASH

MDR report key: 1020836 · Received May 30, 2007

Report

Report Number
2954323-2007-09078
Event Type
Malfunction
Date Received
May 30, 2007
Date of Event
February 14, 2007
Report Date
May 30, 2007
Manufacturer
ABBOTT DIABETES CARE INC.
Product Code
LFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

THE METER HAS BEEN CONFIRMED TO EXHIBIT THE MEMORY OVERWRITE MALFUNCTION. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FA16MAY2006 LETTER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THEY OBSERVED AN ERROR 4 MESSAGE ON THEIR FREESTYLE FLASH METER. THE CUSTOMER'S METER WAS RETURNED FOR INVESTIGATION AND TESTING CONFIRMED THE METER TO EXHIBIT THE MEMORY OVERWRITE MALFUNCTION. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FLASH BLOOD GLUCOSE MONITORING SYSTEM LFR ABBOTT DIABETES CARE INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NI