FDA Adverse Event
Malfunction
Summary report: N
MEDISENSE OPTIUM
MDR report key: 1020542
·
Received May 31, 2007
Report
- Report Number
- 2954323-2007-09267
- Event Type
- Malfunction
- Date Received
- May 31, 2007
- Date of Event
- May 4, 2007
- Report Date
- May 31, 2007
- Manufacturer
- ABBOTT DIABETES CARE INC., USA
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THIS IS AN INITIAL REPORT. THE METER HAS BEEN RETURNED AND AN INVESTIGATION IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR MEDISENSE OPTIUM BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 43MG/DL AND 301MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONES SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEDISENSE OPTIUM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC., USA | NA | 41702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |