FDA Adverse Event Malfunction Summary report: N

BD PEGASUS SAFETY CLOSED IV CATHETER SYSTEM

MDR report key: 10204210 · Received June 26, 2020

Report

Report Number
3006948883-2020-00259
Event Type
Malfunction
Date Received
June 26, 2020
Date of Event
June 8, 2020
Report Date
August 14, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
PENDING
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: A PHYSICAL SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BUT BD WAS PROVIDED WITH TWO PHOTOS OF THE ISSUE FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED FOR THE REPORTED LOT, 9232327, AND NO QUALITY ISSUES WERE FOUND DURING PRODUCTION. OUR QUALITY ENGINEER REVIEWED THE PROVIDED PHOTOS AND OBSERVED THAT THE TOP DISK WAS PUSHED INTO THE BODY. NO OTHER DAMAGE OR DEFECTS WERE OBSERVED. BASED OFF THE PROVIDED PHOTO THE ENGINEER WAS ABLE TO VERIFY THE REPORTED FAILURE MODE. UNFORTUNATELY, WITHOUT A PHYSICAL SAMPLE AVAILABLE FOR EVALUATION A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM EXPERIENCED A DEFECTIVE/DAMAGED ADAPTER SEPTUM. PRODUCT DEFECT WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IN THE INTRACARDIAC ICU, WHEN UNPACKING ON 6.8, IT WAS FOUND THAT THE SEPTUM OF THE NEEDLE WAS BROKEN.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM EXPERIENCED A DEFECTIVE/DAMAGED ADAPTER SEPTUM. PRODUCT DEFECT WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IN THE INTRACARDIAC ICU, WHEN UNPACKING ON 6.8, IT WAS FOUND THAT THE SEPTUM OF THE NEEDLE WAS BROKEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668189 BD PEGASUS SAFETY CLOSED IV CATHETER SYSTEM INTERVASCULAR CATHETER FOZ BD (SUZHOU) 9232327

Patients

Seq Age Sex Outcome Treatment
1 Other