FDA Adverse Event Death Summary report: N

CENTURY CCU BED

MDR report key: 1020375 · Received March 28, 2008

Report

Report Number
1824206-2008-00076
Event Type
Death
Date Received
March 28, 2008
Date of Event
March 18, 2008
Report Date
March 18, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

HOSP ALLEGED THAT A PT HAD EXPIRED ON THE BED AND WHEN THEY FOUND THE PT, THE PT'S HEAD WAS BETWEEN THE MATTRESS AND THE SIDERAIL. (ZONE 3 ENTRAPMENT) THE MATTRESS ON THE BED AT THE TIME OF THE INCIDENT WAS NOT A HILL-ROM PROD BUT A SENTECH STAGE 4 MILLENIUM 3. WHEN THE HILL-ROM TECH ARRIVED AT THE FACILITY, HE FOUND THAT THE SENTECH AIR MATTRESS, WHICH WAS RENTED THRU (B)(4), WAS ALREADY PICKED UP AND NOT ON THE BED. HE INSPECTED THE BED INCLUDING SIDE RAILS AND FOUND EVERYTHING TO BE IN GOOD WORKING CONDITION. THE ACCOUNT REFUSED TO SUPPLY ANY MORE INFO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY CCU BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 896 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Death