FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP

MDR report key: 10201154 · Received June 26, 2020

Report

Report Number
1920898-2020-00755
Event Type
Malfunction
Date Received
June 26, 2020
Date of Event
June 5, 2020
Report Date
July 1, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: 1 SAMPLE WAS RETURNED. BDJ CONFIRMED THAT THE NEEDLE SHIELD WITH HUB WAS DETACHED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231316. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200842714] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. CAPA1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE."

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP NEEDLE SEPARATED FROM THE HUB. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A NEEDLE WITH ITS SHIELD WAS SEPARATED FROM THE HUB.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP NEEDLE SEPARATED FROM THE HUB. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A NEEDLE WITH ITS SHIELD WAS SEPARATED FROM THE HUB.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668064 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP SYRINGE FMF BD MEDICAL - DIABETES CARE 9231316

Patients

Seq Age Sex Outcome Treatment
1 Other