RDCE FRCPS W/ PTS BRD
Report
- Report Number
- 1020279-2020-02752
- Event Type
- Malfunction
- Date Received
- June 25, 2020
- Date of Event
- June 3, 2020
- Report Date
- March 15, 2021
- Manufacturer
- SMITH & NEPHEW, INC.
- Product Code
- JDG
- PMA / PMN Number
- K123598
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE, INTENDED FOR USE IN TREATMENT, WAS RETURNED FOR EVALUATION. THE VISUAL INSPECTION OF THE RETURNED FORCEP CONFIRMS THE TIP IS BENT. THIS DEVICE WAS MANUFACTURED IN 2019. THIS DEVICE SHOWS SIGNS OF SIGNIFICANT WEAR AND USAGE. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE MANUFACTURING RECORDS DID NOT REVEAL A MANUFACTURING ABNORMALITY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. THIS DEVICE IS A REUSABLE INSTRUMENTS THAT CAN BE EXPOSED TO NUMEROUS SURGERIES; DAMAGE FROM REPEATED USE CAN OCCUR. EXPECTED WEAR/TEAR IS LIKELY THE PROBABLE CAUSE OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINTS WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THESE COMPLAINTS; HOWEVER WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.
IT WAS REPORTED THAT AFTER THE PROCEDURE, DURING AN E FIP INSPECTION IT WAS FOUND THAT THE TIP OF THE RDCE FRCPS W/ PTS BRD WAS BENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 658369 | RDCE FRCPS W/ PTS BRD | PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL | JDG | SMITH & NEPHEW, INC. | 19LTCD153 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |