FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK TIP

MDR report key: 10197028 · Received June 25, 2020

Report

Report Number
1213809-2020-00419
Event Type
Malfunction
Date Received
June 25, 2020
Date of Event
June 4, 2020
Report Date
July 24, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096283
PMA / PMN Number
K941562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: EIGHT PHOTOS AND EIGHT LOOSE 1ML SYRINGES WERE RECEIVED AND EVALUATED. IT WAS OBSERVED IN SEVEN PHOTOS AND SEVEN PHYSICAL SAMPLES, THERE WAS A SMALL PERPENDICULAR SCRATCH OUTSIDE THE GRAD LINES APPROXIMATELY 3/16" IN LENGTH AT THE 0.14ML MARKING. ONE SYRINGE HAD MINOR SCUFF MARKS OUTSIDE THE GRAD LINES. IN ALL EIGHT SYRINGES THE SCRATCHES WERE ACCEPTABLE PER PRODUCT SPECIFICATION. DHR WAS PERFORMED. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 9232833 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. A POTENTIAL ROOT CAUSE FOR THE SCRATCHED BARREL DEFECT IS LIKELY ASSOCIATED WITH THE ASSEMBLY PROCESS. SINCE THE DEFECTS OBSERVED ARE WITHIN THE ACCEPTABLE LIMITS, NO CORRECTIVE ACTIONS ARE NECESSARY AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT SCRATCHES WERE FOUND ON THE BD SYRINGE LUER-LOK¿ TIP DURING USE AT THE "0.13-0.15 MARK". THIS OCCURRED WITH 8 SEPARATE SYRINGES, BUT THE DATES ARE UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SCRATCHES OBSERVED AT THE 0.13-0.15 MARK. SIGNIFICANT SCRATCHES OBSERVED AT THE 0.13-0.15 MARK DURING COMPOUNDING. THESE SYRINGES WERE REJECTED FOR USE."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SCRATCHES WERE FOUND ON THE BD SYRINGE LUER-LOK¿ TIP DURING USE AT THE "0.13-0.15 MARK." THIS OCCURRED WITH 8 SEPARATE SYRINGES, BUT THE DATES ARE UNKNOWN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SCRATCHES OBSERVED AT THE 0.13-0.15 MARK. SIGNIFICANT SCRATCHES OBSERVED AT THE 0.13-0.15 MARK DURING COMPOUNDING. THESE SYRINGES WERE REJECTED FOR USE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
657537 BD SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 309628 9232833 30382903096283

Patients

Seq Age Sex Outcome Treatment
1 Other