RADIAL JAW 4 JUMBO BIOPSY FORCEPS
Report
- Report Number
- 3005099803-2008-00295
- Event Type
- Injury
- Date Received
- March 25, 2008
- Date of Event
- February 27, 2008
- Report Date
- February 28, 2008
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- KNW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE COMPLAINANT INDICATED THAT THE DEVICE HAS BEEN DISCARDED AND WILL NOT BE RETURNED FOR EVAL. A DEVICE ANALYSIS CANNOT BE PERFORMED; THEREFORE, THE RELATIONSHIP BETWEEN THE DEVICE AND THE CAUSE OF THE REPORTED EVENT IS UNDETERMINED. THE LOT # WAS NOT PROVIDED BY THE COMPLAINANT; THEREFORE, THE CUSTOMER'S SHIPMENT HISTORY WAS REVIEWED. THIS IDENTIFIED THREE LOTS (9830231, 11319159 AND 11179603) THAT WERE SHIPPED TO THE CUSTOMER PRIOR TO THE EVENT DATE. A REVIEW OF THE DEVICE HISTORY RECORD FOR EACH LOT WAS PERFORMED; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED ONE ADD'L COMPLAINT WAS REPORTED FOR LOT 9830231 FOR A DIFFERENT FAILURE MODE (BROKEN PULL WIRES). NO ADD'L COMPLAINTS WERE REPORTED FOR LOTS 11319159 AND 11179603. THE 2008 15-MO RADIAL JAW 4 BIOPSY FORCEPS PROD FAMILY COMPLAINT TREND REPORT, INCLUSIVE OF ALL FAILURE MODES, WAS REVIEWED; NO UNFAVORABLE TREND WAS NOTED.
A RADIAL JAW 4 JUMBO BIOPSY FORCEPS DEVICE WAS USED IN AN SOPHAGOGASTRODUODENOSCOPY PROCEDURE ON A FEMALE PT IN 2008. ACCORDING TO THE COMPLAINANT, "AFTER A BIOPSY WAS TAKEN FROM THE DUODENUM, THE PT BLED A LOT. THE PHYSICIAN (STATED THAT) THIS PROD WAS TOO BIG AND TOOK TOO LARGE A BITE. THE WRONG ONE WAS TAKEN OUT DURING PREPARATION AND THEY FEEL THE LABELING (IS UNCLEAR) BETWEEN THE JUMBO AND LARGE PACKAGES (AND) THAT (THIS) CAUSED THE ERROR." IT WAS FURTHER REPORTED THAT: "THE PT IS IN INTENSIVE CARE AND HAS REC'D 7 UNITS OF BLOOD..." NO ADD'L INFO HAS BEEN MADE AVAILABLE TO BOSTON SCIENTIFIC CONCERNING THE PT'S PRESENT CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIAL JAW 4 JUMBO BIOPSY FORCEPS | KNW | BOSTON SCIENTIFIC CORPORATION | M00513372 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization| R |