ANATOMICAL SHOULDER DOMELOCK, HUMERAL HEAD, 48-17, R=26.8MM
Report
- Report Number
- 0009613350-2020-00272
- Event Type
- Injury
- Date Received
- June 23, 2020
- Date of Event
- May 11, 2020
- Report Date
- November 4, 2020
- Manufacturer
- ZIMMER SWITZERLAND MANUFACTURING GMBH
- Product Code
- HSD
- UDI-DI
- 00889024485280
- PMA / PMN Number
- K142403
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION WHICH WAS RECEIVED ON JUL 20, 2020. THIS FOLLOW-UP REPORT IS BEING FILLED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE PREVIOUS MEDWATCH. THE MANUFACTURER RECEIVED IMPLANT OPERATIVE AND ARTHROSCOPY REPORT FOR REVIEW. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE AND / OR THE DEVICE(S) BE RETURNED FOR EVALUATION AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE SUBMITTED. ZIMMER¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).
INVESTIGATION RESULTS WERE MADE AVAILABLE. EVENT DESCRIPTION: IT WAS REPORTED PATIENT INITIALLY WAS IMPLANTED WITH A RIGHT SIDE ANATOMICAL SHOULDER ON (B)(6) 2019. THE PATIENT REPORTED PAIN AND WAS SUSPECTED TO HAVE ATROPHY OF THE DELTOID. THE PATIENT UNDERWENT AN ARTHROSCOPIC PROCEDURE ON (B)(6) 2020 AND WAS FOUND TO HAVE BURSITIS AND POSSIBLE MIGRATED ANCHOR (SEE (B)(4)). SAMPLES WERE TAKEN DURING THIS PROCEDURE. AN LOW-GRADE INFECTION WAS DIAGNOSED AND A 2 STAGE REVISION OF THE IMPLANT WAS PERFORMED ON (B)(6) 2020. REVIEW OF RECEIVED DATA: SURGICAL REPORT: MEDICAL RECORDS WERE PROVIDED AND REVIEWED BY A HEALTH CARE PROFESSIONAL. REVIEW OF THE AVAILABLE RECORDS IDENTIFIED THE FOLLOWING: IMPLANTATION SURGICAL REPORT: MEDICAL HISTORY OF ARTHRITIS. NO COMPLICATIONS NOTED DURING INITIAL PROCEDURE. ARTHROSCOPY SURGICAL REPORT: ARTHROSCOPIC EXPLORATION OF RIGHT SHOULDER, SEVERE BURSITIS REQUIRING BURSECTOMY, CUFF INTACT, IMPLANTS INTACT, HOWEVER ANCHOR (OF UNKNOWN DESIGN CONTROL) TO REPAIR SUBSCAPULARIS WHICH SEEMS TO HAVE MIGRATED. CULTURES TAKEN NO INDICATION OF INFECTION NOTED DURING PROCEDURE. PRODUCT EVALUATION: NO PRODUCT WAS RETURNED; THEREFORE, PRODUCT EVALUATION COULD NOT BE PERFORMED. REVIEW OF PRODUCT DOCUMENTATION: DEVICE PURPOSE: ALL INVOLVED DEVICES ARE INTENDED FOR TREATMENT. PRODUCT COMPATIBILITY: THE COMPATIBILITY CHECK WAS PERFORMED AND SHOWED THAT THE PRODUCT COMBINATION WAS APPROVED BY ZIMMER BIOMET. DHR REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. NCR(S): LOT: 2912819: 1 PART WAS SCRAPPED DUE TO INTERNAL HANDLING ERROR. NO NCR WITH A POTENTIAL CORRELATION TO THE REPORTED EVENT WAS FOUND. STERILIZATION CERTIFICATE: THE GAMMA STERILIZATION SPECIFICATION OF THE DEVICES CERTIFIES THE SUITABILITY OF STERILIZATION. THE IRRADIATION CERTIFICATES OF THE AFFECTED LOT NUMBERS HAVE BEEN REVIEWED AND WAS FOUND TO BE ACCORDING TO SPECIFICATIONS. CONCLUSION: IT WAS REPORTED THAT FOLLOWING ARTHROSCOPIC PROCEDURE (B)(6) 2020, A LOW-GRADE INFECTION WAS DIAGNOSED AND A 2 STAGE REVISION OF THE IMPLANT WAS PERFORMED ON (B)(6) 2020. THE QUALITY RECORDS SHOW THAT ALL SPECIFIED CHARACTERISTICS HAVE MET THE SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. THEREFORE, THE INVESTIGATION DID NOT IDENTIFY A NONCONFORMANCE OR A COMPLAINT OUT OF BOX (COOB). STERILIZATION CERTIFICATIONS WERE REVIEWED AND FOUND TO BE CONFORMING WITH NO APPLICABLE DEVIATIONS. DEVICES WERE VERIFIED TO HAVE GONE THROUGH ACCEPTABLE STERILIZATION PROCESS FOLLOWING ISO/AAMI/ASTM & EU PUBLISHED GUIDELINES. THERE ARE MULTIPLE FACTORS THAT MAY CONTRIBUTE TO AN INFECTION THAT ARE OUTSIDE THE CONTROL OF ZIMMER BIOMET, SUCH AS EXTERNAL FACTORS, I.E. HOSPITAL/SURGICAL ENVIRONMENT, PROVIDER RELATED RISK FACTORS, AND/OR PATIENT COMORBIDITIES/RISK FACTORS. AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUES IDENTIFIED AFFECTING IMPLANT SAFETY OR EFFECTIVENESS, THEREFORE IMPLANTED PRODUCTS ARE NOT IDENTIFIED AS THE SOURCE OR CONTRIBUTING TO THE REPORTED INFECTION. BASED ON THE GIVEN INFORMATION AND THE RESULTS OF THE INVESTIGATION, WE WERE NOT ABLE TO IDENTIFY A SPECIFIC ROOT CAUSE FOR THE INFECTION. THE NEED FOR CORRECTIVE MEASURES IS NOT INDICATED AND ZIMMER SWITZERLAND MANUFACTURING GMBH CONSIDERS THIS CASE AS CLOSED. ZIMMER BIOMET'S REFERENCE NUMBER OF THIS FILE IS (B)(4). THE FOLLOWING REPORTS ARE ASSOCIATED WITH THIS EVENT: 0009613350-2020-00274-3; 0009613350-2020-00286-3.
ADDITIONAL INFORMATION WHICH WAS RECEIVED ON SEP 15, 2020 THIS FOLLOW-UP REPORT IS BEING FILLED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE PREVIOUS MEDWATCH. THE MANUFACTURER DID NOT RECEIVE ANY DOCUMENTS FOR REVIEW. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE AND / OR THE DEVICE(S) BE RETURNED FOR EVALUATION AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE SUBMITTED. ZIMMER¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).
IT WAS REPORTED THAT FOLLOWING ARTHROSCOPIC PROCEDURE, INFECTION WAS DIAGNOSED AND A 2 STAGE REVISION OF THE IMPLANT HAS BEEN PLANNED.
INVESTIGATION HAS BEEN COMPLETED.
PATIENT WAS IMPLANTED ON THE RIGHT SIDE AND EXPERIENCED PAIN 6 MONTHS POST THE SURGERY. PATIENT UNDERWENT AN ARTHROSCOPIC PROCEDURE AND WAS FOUND TO HAVE BURSITIS AND POSSIBLE MIGRATED ANCHOR. IT WAS REPORTED THAT FOLLOWING ARTHROSCOPIC PROCEDURE INFECTION WAS DIAGNOSED AND A 2 STAGE REVISION OF THE IMPLANT HAS BEEN PERFORMED.
CONCOMITANT MEDICAL PRODUCT: MD HYBRID GLENOID BASE 4MM D BASE 4MM; CATALOG#: 113954; LOT#: 945140. THERAPY DATE: UNKNOWN. THE MANUFACTURER RECEIVED OTHER SOURCE DOCUMENTS FOR REVIEW. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. (B)(4).
PATIENT WAS IMPLANTED ON THE RIGHT SIDE AND EXPERIENCED INFECTION A REVISION SURGERY HAS BEEN PLANNED. THIS REVISION WILL TAKE PLACE IN 2 STAGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 649323 | ANATOMICAL SHOULDER DOMELOCK, HUMERAL HEAD, 48-17, R=26.8MM | ANATOMICAL SHOULDER DOMELOCK SYSTEM | HSD | ZIMMER SWITZERLAND MANUFACTURING GMBH | N/A | 2867867 | 00889024485280 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization | SEE H10 NARRATIVE. |