FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
MDR report key: 10171885
·
Received June 18, 2020
Report
- Report Number
- 3013756811-2020-61673
- Event Type
- Malfunction
- Date Received
- June 18, 2020
- Date of Event
- May 28, 2020
- Report Date
- June 18, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00085006613373
- PMA / PMN Number
- P180008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CARTRIDGE USER GUIDE CAUTIONS THAT INSULIN SHOULD BE AT ROOM TEMPERATURE BEFORE FILLING A CARTRIDGE. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. REPORTEDLY, THE CUSTOMER LOADED COLD INSULIN INTO THE CARTRIDGE. CUSTOMER¿S BLOOD GLUCOSE RANGED FROM 110-170 MG/DL. REPORTEDLY, THE CUSTOMER CONTINUED USING THE EXISTING CARTRIDGE FOR INSULIN THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 631991 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1002717 | M508319 | 00085006613373 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |