FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 10165161 · Received June 17, 2020

Report

Report Number
3004753838-2020-059398
Event Type
Malfunction
Date Received
June 17, 2020
Date of Event
June 1, 2020
Report Date
August 19, 2020
Manufacturer
DEXCOM, INC.
Product Code
QBJ
UDI-DI
00386270000385
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A TRANSMITTER FAILED ERROR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND FAILED DUE TO 0V. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND TRANSMITTER FAILED ERROR WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A TRANSMITTER FAILED ERROR OCCURRED. DATA WAS EVALUATED AND THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
627851 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-02 5268156 00386270000385

Patients

Seq Age Sex Outcome Treatment
1 61 YR