FDA Adverse Event Malfunction Summary report: N

BD HOLDER ONE USE

MDR report key: 10164152 · Received June 17, 2020

Report

Report Number
2243072-2020-00919
Event Type
Malfunction
Date Received
June 17, 2020
Date of Event
May 28, 2020
Report Date
August 26, 2020
Manufacturer
BECTON DICKINSON
Product Code
JKA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 7/3/2020. H.6. INVESTIGATION: BD RECEIVED 7 SAMPLES FROM THE CUSTOMER FOR INVESTIGATION. ALL SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR A MOLDING DEFECT WITH THE INCIDENT LOT WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SHARP MOLDING DEFECT WAS DISCOVERED ON WING WITH A BD HOLDER ONE USE. THIS OCCURRED ON 30 SEPARATE OCCASIONS PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: IT HAS BEEN REPORTED THAT A BURR ON WING IS DANGEROUS BECAUSE IT TOUCHES PATIENT¿S SKIN DURING BLOOD COLLECTION. OF HOSPITAL STOCKS, SOME (20-30) HAVE A BURR ON WING.

Additional Manufacturer Narrative · 1

FOR OEM MANUFACTURING SITES: IN THIS MDR, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AS (B)(4) IS AN OEM MANUFACTURING SITE. MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A SHARP MOLDING DEFECT WAS DISCOVERED ON WING WITH A BD HOLDER ONE USE. THIS OCCURRED ON 30 SEPARATE OCCASIONS PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: IT HAS BEEN REPORTED THAT A BURR ON WING IS DANGEROUS BECAUSE IT TOUCHES PATIENT¿S SKIN DURING BLOOD COLLECTION. OF HOSPITAL STOCKS, SOME (20-30) HAVE A BURR ON WING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
629689 BD HOLDER ONE USE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON DICKINSON 2002BD001

Patients

Seq Age Sex Outcome Treatment
1 Other