FDA Adverse Event Other Summary report: N

MICROTARGETING DRIVE SYSTEM

MDR report key: 1016384 · Received March 12, 2008

Report

Report Number
3005677147-2008-00001
Event Type
Other
Date Received
March 12, 2008
Date of Event
February 12, 2008
Report Date
March 12, 2008
Manufacturer
FHC, INC.
Product Code
HAW
PMA / PMN Number
K011775
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INFO REGARDING THIS EVENT WAS INITIALLY RECEIVED FROM THE DISTRIBUTOR'S REPRESENTATIVE. FHC EVALUATED THE MICROTARGETING PLATFORM FOR POSSIBLE TARGETING ERROR AND CONCLUDED THAT THE DEVICE WAS MANUFACTURED AND PERFORMED ACCORDING TO SPECIFICATIONS. FURTHER INVESTIGATION (IE, PHONE CONVERSATION WITH NEUROSURGEON) INTO THE INCIDENT CONCLUDED THAT THERE WAS NO FAILURE TO ANY FHC PRODUCT IN THIS SURGERY, WHICH INCLUDED THE MICROTARGETING DRIVE AND ENCODER, MICROTARGETING ELECTRODES AND THE AXON GUIDELINE SYSTEM 3000. IN ADDITION, ALL FHC PRODUCTS NOTED HERE WERE USED IN THE SUBSEQUENT SURGERY AND FUNCTIONED PROPERLY. NO PT INJURY OR HARM WAS NOTED DURING THIS CASE OTHER THAN A SECOND SURGERY TO REPOSITION THE DBS LEAD. NOTE: THE DBS LEAD IS NOT AN FHC PRODUCT AND THIS REPORT IS BEING FILED TO DOCUMENT THE USE OF THE FHC PRODUCTS DURING THE EVENT.

Description of Event or Problem · 1

THE DISTRIBUTOR'S REPRESENTATIVE REPORTED THAT FOLLOWING BILATERAL LEAD PLACEMENT (DATE WAS NOT PROVIDED); LEAD MIGRATION HAD BEEN SUSPECTED BY THE SURGEON. NO PT SYMPTOMS WERE REPORTED. THE PT UNDERWENT A SECOND SURGERY (DATE UNK), WHERE THE LEAD WAS FOUND SECURE, BUT HAD BEEN PLACED IN AN AREA ESTIMATED TO BE 1.5 TO 2 CM HIGHER THAN THE ORIGINAL TARGET LOCATION. THERE HAD BEEN NO MALFUNCTION OF THE LEAD; THE DEPTH CAP HAD BEEN PROPERLY INSTALLED AND THE LEAD HAD BEEN MEASURED PRIOR TO PLACEMENT. DURING THE REVISION PROCEDURE, THE LEAD WAS REPOSITIONED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICROTARGETING DRIVE SYSTEM STEREOTACTIC POSITIONER HAW FHC, INC. 66-MP-BT-NNN

Patients

Seq Age Sex Outcome Treatment
1 Other MICROTARGETING DRIVE WITH ENCODER OPTION| MICROTARGETING ELECTRODES| AXON GUIDELINE SYSTEM 3000