FDA Adverse Event Other Summary report: N

EXH VLV ASSY, PED, MILL, ENGLISH

MDR report key: 1016373 · Received March 13, 2008

Report

Report Number
2020676-2008-00003
Event Type
Other
Date Received
March 13, 2008
Date of Event
February 4, 2008
Report Date
March 13, 2008
Manufacturer
SECHRIST INDUSTRIES, INC.
Product Code
CBP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

APPEARS TO BE USER ERROR. FOLLOW UP REPORT WILL BE SUBMITTED FOR EVALUATION.

Description of Event or Problem · 1

DISTRIBUTOR IN (B) (4) REPORTED THAT CUSTOMER AT HOSPITAL APPARENTLY ATTEMPTED TO USE VENTILATOR WITH REFERENCE PRESSURE LINE STUCK INTO A THREADED HOLE IN THE FRONT OF THE PEDIATRIC EXHALATION BLOCK. THE PEDIATRIC EXHALATION BLOCK IS USED TO VENTILATE PTS RANGING FROM 10 LBS TO 100 LBS, NOT NEONATES OR PREMATURE INFANTS. THE THREADED HOLE IS PLUGGED WITH AT LEAST TWO SETSCREWS TO PREVENT FULL INSERTION OF THE SENSE LINE. THE REFERENCE PRESSURE CONNECTOR IS DESIGNED TO BE A TAPERED SLIP FIT INTO A SMOOTH SIDE TAPERED HOLE IN THE LOWER FRONT OF THE BLOCK. THE BLOCK LABEL CLEARLY IDENTIFIES THE HOLES THE PROXIMAL LINE, REFERENCE PRESSURE LINE AND "FROM PT" HOSE CONNECTIONS ARE TO BE MADE INTO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXH VLV ASSY, PED, MILL, ENGLISH PEDIATRIC EXHALATION VALVE CBP SECHRIST INDUSTRIES, INC. 21442-1

Patients

Seq Age Sex Outcome Treatment