FDA Adverse Event Malfunction Summary report: N

GCX WALL BRACKET, VARIABLE HEIGHT

MDR report key: 1016198 · Received March 18, 2008

Report

Report Number
1220063-2008-00008
Event Type
Malfunction
Date Received
March 18, 2008
Date of Event
February 14, 2008
Report Date
March 17, 2008
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC. (IT/M)
Product Code
DRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS REPORTED, THE SCREWS FROM THE GCX MOUNTING BRACKET CAME LOOSE CAUSING DRAEGER MONITOR TO FALL FORWARD. THE GCX MOUNTING BRACKET IS A NON DRAEGER PRODUCT. THE DRAEGER DEVICE DID NOT CAUSE THIS EVENT. WE HAVE NOTIFIED THE MOUNTING BRACKET MFR (GCX CORP) OF THIS EVENT FOR FURTHER INVESTIGATION AND POSSIBLE CORRECTIVE ACTION BY GCX CORP. DRAEGER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TREATMENT OF A PT A NURSE TOUCHED THE DRAEGER MONITOR WHICH FELL FORWARD AND HIT HER IN THE FACE. SHE SUFFERED MINOR CUTS AND BRUISING AND HAD TO BE RELIEVED OF HER DUTIES. A FAULT WAS DIAGNOSED WITH GCX BRACKET DUE TO THE TENSION BOLTS WORKING LOOSE CAUSING THE 5 INCH PLUNGE PLATE TENSIONING MECHANISM TO BECOME LOOSE AND ALLOWING THE DRAEGER MONITOR TO FALL. THE DRAEGER MONITOR WAS MOUNTED ON THE GCX BRACKET. THE NURSE RETURNED TO DUTY THE FOLLOWING DAY. THE BRACKET WAS TENSIONED ON INSTILLATION, BUT SUBSEQUENTLY WORKED LOOSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GCX WALL BRACKET, VARIABLE HEIGHT PATIENT MONITOR MONITORING BRACKET DRT DRAEGER MEDICAL SYSTEMS, INC. (IT/M) S1-0021-19C P030357

Patients

Seq Age Sex Outcome Treatment
1