FDA Adverse Event Malfunction Summary report: N

BD SAFE-CLIP NEEDLE CLIPPER

MDR report key: 10160471 · Received June 16, 2020

Report

Report Number
2243072-2020-00918
Event Type
Malfunction
Date Received
June 16, 2020
Date of Event
June 8, 2020
Report Date
June 8, 2020
Manufacturer
BECTON DICKINSON
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE EXPIRATION DATE: NA. OEM MANUFACTURE: THE MANUFACTURING LOCATION FOR THIS PRODUCT IS (B)(4). THIS SITE IS AN OEM MANUFACTURING SITE. THEREFORE, BD CORPORATE HEADQUARTERS IN (B)(4) HAS BEEN LISTED AND THE (B)(4) FDA REGISTRATION NUMBER HAS BEEN USED FOR THE MANUFACTURE REPORT NUMBER. INVESTIGATION SUMMARY: LEVEL A INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR SAFE CLIP NOT CLIPPING (NO LONGER WORKS) ON LOT # 9059012. A REVIEW OF RISK MANAGEMENT (B)(4) INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (SAFECLIP, NOT CLIPPING (NO LONGER WORKS) WAS CAPTURED AND ADDRESSED APPROPRIATELY. INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. ACCORDING WITH THE DHR REVIEW INFORMATION, THE PROBLEM ¿NO CLIPPING¿ HAS NO NHB ASSEMBLY PROCESS RELATION, ALL SAMPLES OF SAFE CLIP DISPOSABLE CUTTER (USA) PASSED FUNCTIONAL TEST (SAMPLE PLAN C=0 AND AQL=1). INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD SAFE-CLIP¿ NEEDLE CLIPPERS WERE UNABLE TO SLIP/JAMMED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: A PATIENT / CUSTOMER BOUGHT BD SAFE CLIP NEEDLE CUTTER IN FEBRUARY. UNFORTUNATELY IT NO LONGER WORKS!

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
624595 BD SAFE-CLIP NEEDLE CLIPPER NEEDLE CLIPPER FMI BECTON DICKINSON 9059012

Patients

Seq Age Sex Outcome Treatment
1 Other