FDA Adverse Event Injury Summary report: N

UNIVERSAL GLENOID - BASEPLATE SMALL

MDR report key: 10158180 · Received June 16, 2020

Report

Report Number
1220246-2020-01893
Event Type
Injury
Date Received
June 16, 2020
Date of Event
May 18, 2020
Report Date
June 16, 2020
Manufacturer
ARTHREX, INC.
Product Code
PHX
UDI-DI
00888867057340
PMA / PMN Number
K142863
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON HAS STATED THAT THE POSITION OF THE IMPLANTED BASE PLATE IS NOT IN A PROPER POSITION AND THEREFORE THE SURGEON DECIDED TO PERFORM A REVISION SURGERY. DURING THE REVISION SURGERY THE SURGEON DECIDED TO MAKE AN EASIER ACCESS TO THE GLENOID TO REPLACE THE WHOLE PROSTHESES. IT WAS FURTHER REPORTED THAT IT WAS FOUND ON THE X-RAY THAT DUE TO THE POSITION OF THE BASE PLATE A SCAPULAR NOTCHING WAS CAUSED. THE FOLLOWING IMPLANTS WERE EXPLANTED: BASE PLATE AR-9500-01 (PART NUMBER UNKNOWN AS THE PROVIDED ONE IS INCORRECT), GLENOSPHERE AR-9504S-04, STOCK AR-9501-08CPC, 2X PERIPHERE SCREWS (LOCKING) (PART NUMBER UNKNOWN), CENTRAL SCREWS (LOCKING) (PART NUMBER UNKNOWN), PE-INLAY (PART NUMBER UNKNOWN). UPDATE 07-JUN-2020: FURTHER INFORMATION WERE PROVIDED THAT THE BASE PLATE WAS NOT OPTIMALLY PLACED AND THE PLATE DID NOT SINK. DURING THE SURGERY IT WAS ALSO DISCOVERED THAT THE PLATE HAD GOOD OSSEOINTEGRATION. THE SIZE THE CENTRAL SCREW IS AN AR-9165-15.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
621999 UNIVERSAL GLENOID - BASEPLATE SMALL SHOULDER PROSTHESIS, REVERSE CONFIGURATION PHX ARTHREX, INC. UNIVERSAL GLENOID - BASEPLATE SMALL UNK 00888867057340

Patients

Seq Age Sex Outcome Treatment
1 Other