HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00352
- Event Type
- Injury
- Date Received
- June 16, 2020
- Report Date
- May 20, 2020
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PATIENT INFORMATION WAS NOT PROVIDED. SERIAL NUMBER IS UNKNOWN. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. AS THE SERIAL NUMBER IS UNKNOWN, THE DEVICE MANUFACTURE DATE COULD NOT BE DETERMINED. THIS INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF MADE AVAILABLE. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T. THE INCIDENT OCCURRED IN (B)(6). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THROUGH LITERATURE REVIEW LIVANOVA LEARNED ABOUT TWO (2) MORE PATIENTS INFECTED WITH M. CHIMAERA AT (B)(6). WITHIN THE ARTICLE IT STATED THAT BOTH PATIENTS HAD UNDERGONE CARDIAC SURGERIES BEFORE (B)(6) 2014 AND HEATER-COOLER SYSTEMS 3T WERE USED. THE PATIENTS WERE SUBJECTED TO TREATMENT WITH ANTIBIOTICS THERAPY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 620920 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |