MDR report key: 10148050 · Received June 12, 2020

Report

Report Number
3006695864-2020-00296
Event Type
Injury
Date Received
June 12, 2020
Date of Event
May 21, 2020
Report Date
October 29, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
HNO
UDI-DI
05050474573468
PMA / PMN Number
K060372
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE (B)(4) AND CAPA-010215.

Additional Manufacturer Narrative · 1

(B)(4). A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, SERVICE HISTORY, AND TRENDING WAS REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. A LABELING REVIEW WAS CONDUCTED; THE OPERATOR MANUAL FOR THE SYSTEM WAS REVIEWED AND FOUND TO INCLUDE ADEQUATE INSTRUCTIONS FOR USE, WARNINGS AND OPERATIONAL ERRORS. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR FEMTOSECOND LASER SYSTEM SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. JOHNSON & JOHNSON SURGICAL VISION WILL CONTINUE TO MONITOR THIS TYPE OF COMPLAINTS. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A LASER VISION CORRECTION PATIENT HAD SURGERY ON (B)(6) 2020 AND A VERTICAL GAS BREAK THROUGH (VGBT) WAS NOTICED. THE VGBT WAS NOTICED AT THE START OF THE RASTER PATTERN. THE SURGEON LIFTED THE FLAP AND COMMENTED THAT THE GAS BROKE THROUGH AT A SUTURE SITE FROM THE PREVIOUS PENETRATING KERATOPLASTY PROCEDURE. THE SURGEON WAS CREATING AND LIFTING THE FLAP TO GET A REFACTIVE EFFECT. LOSS OF VISION IS UNKNOWN AT THIS TIME. THERE WAS NO SURGICAL INTERVENTION. A BANDAGE CONTACT LENS (BCL) WAS PLAED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
613584 IFS FEMTOSECOND LASER HNO JOHNSON & JOHNSON SURGICAL VISION, INC. J20007D 05050474573468

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention