FDA Adverse Event Malfunction Summary report: N

2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM

MDR report key: 10146588 · Received June 11, 2020

Report

Report Number
8030965-2020-04124
Event Type
Malfunction
Date Received
June 11, 2020
Report Date
May 31, 2020
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HTW
UDI-DI
07611819166264
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. CUSTOMER QUALITY INVESTIGATION: THE DRILL BIT W/DOUBLE MARKING L140/115 3 WAS NOT RETURNED, AND THE INVESTIGATION WILL BE COMPLETED BASED ON THE SUPPLIED IMAGE FROM THE ATTACHMENTS (¿SOURCE FILE - NOVEDAD DE SERVICIO CL REINA SOFIA¿ IN THE NOTES & ATTACHMENTS SECTION OF THE PRODUCT COMPLAINT). THE IMAGES WERE REVIEWED, AND THE COMPLAINT CONDITION WAS CONFIRMED AS THE IMAGE SHOWED THE ENTIRE DRILLING BIT BROKEN OFF THE SHAFT. AS THE BIT WAS NOT RETURNED AN AS RECEIVED, DIMENSIONAL, AND MATERIAL REVIEW WERE NOT APPLICABLE. MANUFACTURING RECORD EVALUATION. A MANUFACTURING RECORD EVALUATION COULD NOT BE PERFORMED AS THE LOT NUMBER COULD NOT BE DETERMINED FROM THE PROVIDED IMAGE. DOCUMENT/SPECIFICATION REVIEW: DRAWING(S) REVIEWED: (CURRENT & MANUFACTURED REVISIONS). CONCLUSION THE OVERALL COMPLAINT WAS CONFIRMED AS THE DRILL BIT WAS PICTURED BROKEN. NO DEFINITIVE ROOT CAUSE WAS ABLE TO BE DETERMINED AS THE CIRCUMSTANCES SURROUNDING THE EVENT ARE UNKNOWN. DURING THE INVESTIGATION NO UNIDENTIFIED PRODUCT DESIGN/MANUFACTURING ISSUES OR DISCREPANCIES WERE OBSERVED (BASED ON THE IMAGES) THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. (B)(6) . WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020 THAT A 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM BROKE. THE 1.8MM DRILL BIT/QC/100MM WAS ALSO FOUND TO HAVE A BROKEN TIP. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. THIS COMPLAINT INVOLVES THREE (3) DEVICES. THIS REPORT IS FOR ONE (1) 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
610541 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM BIT,DRILL HTW OBERDORF SYNTHES PRODUKTIONS GMBH 07611819166264

Patients

Seq Age Sex Outcome Treatment
1 DRILL BIT Ø1.8 L100/75 2FLUTE| DRILL BIT Ø2.5 L110/85 2FLUTE