FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1014302 · Received March 13, 2008

Report

Report Number
2023826-2008-00370
Event Type
Malfunction
Date Received
March 13, 2008
Date of Event
February 20, 2008
Report Date
February 21, 2008
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P900048
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS- VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC TORN INTO PIECES, WITH A PIECE TORN OFF AND MISSING. ONE HAPTIC WAS TORN OFF. THE OTHER HAPTIC WAS BENT. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION. CONCLUSIONS -AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CANNOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED AN AQ2003V SILICONE THREE PIECE LENS AND THE LENS TORE, THE LENS WAS CUT TO REMOVE WITH NO PATIENT INJURY, NO ENLARGED INCISION OR SUTURES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY AQ2003V NA

Patients

Seq Age Sex Outcome Treatment
1 INJECTOR MODEL MSI-PM - LOT NUMBER UNK| VISCOELASTIC MODEL STAARVISC II-LOT NUMBER UNK| CARTRIDGE MODEL AQ CARTRIDGE-FP-LOT NUMBER 1234221