FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 1014295 · Received March 12, 2008

Report

Report Number
1720753-2008-17938
Event Type
Malfunction
Date Received
March 12, 2008
Date of Event
March 3, 2008
Report Date
March 11, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC.
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE POWER CONTROL AND SURGE SUPPRESSOR PCBS AND THE POWER SWITCH. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE WORKSTATION ON THE 2800 SYSTEM IS NOT BOOTING UP. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC. 2800 NA

Patients

Seq Age Sex Outcome Treatment
1