FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 1014295
·
Received March 12, 2008
Report
- Report Number
- 1720753-2008-17938
- Event Type
- Malfunction
- Date Received
- March 12, 2008
- Date of Event
- March 3, 2008
- Report Date
- March 11, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE POWER CONTROL AND SURGE SUPPRESSOR PCBS AND THE POWER SWITCH. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE WORKSTATION ON THE 2800 SYSTEM IS NOT BOOTING UP. THERE WAS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | 2800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |