FDA Adverse Event Malfunction Summary report: N

ADVIA CENTAUR

MDR report key: 1014027 · Received March 14, 2008

Report

Report Number
2432235-2008-00007
Event Type
Malfunction
Date Received
March 14, 2008
Date of Event
January 20, 2008
Report Date
January 21, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD
Product Code
MMI
PMA / PMN Number
K053020
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE HIGH VALUES REPORTED ON PATIENT SAMPLES AND QC VALUES WERE DUE TO THE EMPTY WASH 1 RESERVOIR. A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE ACCOUNT AND CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

SEVERAL FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON A PATIENTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE ORIGINAL SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE HIGH VALUES REPORTED ON PATIENT SAMPLES, AND QC VALUES WERE DUE TO THE EMPTY WASH 1 RESERVOIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVIA CENTAUR IMMUNO-ASSAY ANALYZER MMI SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD NA NA

Patients

Seq Age Sex Outcome Treatment
1