ADVIA CENTAUR
Report
- Report Number
- 2432235-2008-00007
- Event Type
- Malfunction
- Date Received
- March 14, 2008
- Date of Event
- January 20, 2008
- Report Date
- January 21, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD
- Product Code
- MMI
- PMA / PMN Number
- K053020
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE HIGH VALUES REPORTED ON PATIENT SAMPLES AND QC VALUES WERE DUE TO THE EMPTY WASH 1 RESERVOIR. A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE ACCOUNT AND CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
SEVERAL FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON A PATIENTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE ORIGINAL SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE HIGH VALUES REPORTED ON PATIENT SAMPLES, AND QC VALUES WERE DUE TO THE EMPTY WASH 1 RESERVOIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVIA CENTAUR | IMMUNO-ASSAY ANALYZER | MMI | SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |