ADVIA CENTAUR
Report
- Report Number
- 2432235-2008-00023
- Event Type
- Malfunction
- Date Received
- March 14, 2008
- Date of Event
- February 8, 2008
- Report Date
- February 8, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD
- Product Code
- MMI
- PMA / PMN Number
- P030029
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED ASSAY RESULTS WAS AN EMPTY WASH 1 CONTAINER DUE TO WASH 1 SENSOR FAILURE. A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE ACCOUNT AND CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
FALSELY ELEVATED ASSAY RESULTS WERE OBTAINED ON SEVERAL PATIENT SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIANS. THE SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED ASSAY RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED RESULTS WAS AN EMPTY WASH 1 CONTAINER DUE TO WASH 1 SENSOR FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVIA CENTAUR | IMMUNO-ASSAY ANALYZER | MMI | SIEMENS HEALTHCARE DIAGNOSTICS MFG LTD | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |