FDA Adverse Event Malfunction Summary report: N

PUMP MMT-512LNAP PRDGM INS PL EN US LN

MDR report key: 1014016 · Received March 13, 2008

Report

Report Number
2032227-2008-00479
Event Type
Malfunction
Date Received
March 13, 2008
Date of Event
February 18, 2008
Report Date
February 18, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K030531
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO AUDIO TONE DUE TO A BROKEN TRANSDUCER WIRE ON THE INTERFACE BOARD. THE INSULIN PUMP HAD A NOISY MOTOR DURING THE REWIND. THE PROBLEM WAS ISOLATED TO THE MOTOR. NO ERROR ALARMS WERE NOTED.

Description of Event or Problem · 1

THE CUSTOMER'S MOTHER CALLED TO REPORT AN ERROR ALARM ON THE INSULIN PUMP. THE MOTHER THEN STATED THAT THE INSULIN PUMP NO LONGER HAS AUDIO TONES. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP DID NOT BEEP DURING THE TONE TEST. THE MOTHER ALSO REPORTED A BLOOD GLUCOSE READING OF 300 MG/DL, WHICH THE CUSTOMER TREATED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-512LNAP PRDGM INS PL EN US LN INFUSION PUMP LZG MEDTRONIC MINIMED MMT-512LNAP