LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1014000
·
Received March 13, 2008
Report
- Report Number
- 3015876-2008-00175
- Event Type
- Malfunction
- Date Received
- March 13, 2008
- Date of Event
- January 24, 2008
- Report Date
- February 15, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE WOULD NOT POWER UP ON AC POWER. THE POWER SUPPLY ASSEMBLY WAS REPLACED, THEN PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
FOUND DURING DAILY TESTING. ACCORDING TO THE REPORTER, THE DEVICE HAD NO AC OPERATION. NO PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.