INRATIO
Report
- Report Number
- 2954730-2008-00133
- Event Type
- Malfunction
- Date Received
- March 14, 2008
- Date of Event
- February 14, 2000
- Report Date
- March 14, 2008
- Manufacturer
- HEMOSENSE SJ
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DISCREPANT RESULTS ( ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: 2008, INRATIO 5.0, LAB 5.2, MEAN 5.1, CONFIDENCE LIMITS CANNOT BE DETERMINED. DATE: SAME DAY, INRATIO 3.3, LAB 5.2, MEAN 4.25, CONFIDENCE LIMITS: 2.4-6.1. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR DATE SET # 1, THE READINGS ARE CONSIDERED ACCURATE BASED ON "AREA OUTSIDE THE ACCEPTANCE REGION" TABLE. FOR DATA SET # 2, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME.
CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE: 2008, INRATIO: 5.0, LAB: 5.2; DATE: SAME DAY, INRATIO: 3.3, LAB: 5.2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE SJ | BOXED METER, ENGLISH | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |