FDA Adverse Event Injury Summary report: N

BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 36 SIZE M

MDR report key: 10136927 · Received June 10, 2020

Report

Report Number
3005180920-2020-00324
Event Type
Injury
Date Received
June 10, 2020
Date of Event
June 5, 2020
Report Date
June 10, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809200
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 25 MAY 2020: LOT 1900806: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-MAY-2019. EXPIRATION DATE: 2024-04-23. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH NO OTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING INSTABILITY 7 MONTHS AFTER THE PRIMARY SURGERY WHICH WAS CAUSED BY A LEG LENGTH DISCREPANCY. THE PRIMARY SURGERY WAS NOT AMIS. DR. (B)(6) USES ORTHOGRID ON HIS C-ARM FLUOROSCOPY AND FELT THAT THE LENGTH AND OFFSET LOOKED GOOD DURING THE CASE. HE SAID THE POST-OP X-RAYS LOOKED GOOD AS WELL, BUT THE PATIENT EXPERIENCED A "SHORTENED" FEELING OF THE LEG WHEN SHE WAS WEIGHT-BEARING. THE SURGEON REVISED THE HEAD AND IMPLANTED THE OPTION SLEEVE. THE LINER WASN'T REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601781 BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 36 SIZE M CERAMIC BALL HEAD LZO MEDACTA INTERNATIONAL SA 01.29.209 1900806 07630030809200

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention