BD LUER-LOK TIP SYRINGE
Report
- Report Number
- 1911916-2020-00540
- Event Type
- Malfunction
- Date Received
- June 9, 2020
- Date of Event
- May 28, 2020
- Report Date
- June 1, 2020
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). INVESTIGATION SUMMARY: ONE PHOTO WAS PROVIDED FOR EVALUATION. THE PHOTO SHOWS THE PACKAGING BLISTER TOP WEB AND THE BOTTOM PART OF A SYRINGE, THE SYRINGE HAS BEEN USED. IT HAS A TIP CAP. IT HAS A FOREIGN MATTER LIKE HAIR PLASTIC OF ABOUT 3/8¿ LONG. IT HAS ADHERED TO THE INNER WALL OF THE SYRINGE BARREL. EVEN THOUGH THE PRESENCE OF FOREIGN MATTER IS CONFIRMED, THE SOURCE IS UNKNOWN. THE SYRINGE HAS BEEN USED. FURTHERMORE, A PHYSICAL SAMPLE WOULD BE HELPFUL TO FULLY INVESTIGATE AND CONFIRM THE TYPE OF FOREIGN MATTER AND POSSIBLY THE SOURCE. AT THIS TIME NO FURTHER ACTIONS ARE TAKEN. HOWEVER, MONITORING OF THIS LOT WILL BE PERFORMED FOR THIS TYPE OF SYMPTOM. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: THIS IS THE 1ST COMPLAINT FOR LOT # 62310 FOR THIS TYPE OF DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION FOR THIS TYPE OF DEFECT DURING THE ENTIRE PRODUCTION RUN OF THIS BATCH. ROOT CAUSE DESCRIPTION: ROOT CAUSE COULD NOT BE DETERMINED. EVEN THOUGH THE PRESENCE OF FOREIGN MATTER IS CONFIRMED, THE SOURCE IS UNKNOWN. THE SYRINGE HAS BEEN USED. FURTHERMORE, A PHYSICAL SAMPLE WOULD BE HELPFUL TO FULLY INVESTIGATE AND CONFIRM THE REPORTED CONDITION. RATIONALE: CAPA NOT REQUIRED AT THIS TIME.
IT WAS REPORTED THAT A "PLASTIC HAIR" WAS FOUND IN THE BD LUER-LOK¿ TIP SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "A PHARMACIST REPORTED TO HIM THERE WAS A "PLASTIC HAIR" IN SYRINGE 309653 LOT 0062310 FOUND AROUND 5:00 PM ON (B)(6)2020. PHARMACIST TOOK A PHOTO OF IT DID NOT KNOW IF THEY STILL HAD THE SYRINGE OR IF ANYONE WAS HARMED WHEN IT WAS DISCOVERED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 600179 | BD LUER-LOK TIP SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 309653 | 0062310 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |