FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 1013284 · Received March 14, 2008

Report

Report Number
1824206-2008-00053
Event Type
Malfunction
Date Received
March 14, 2008
Date of Event
February 27, 2008
Report Date
February 27, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. THE HILL-ROM TECHNICIAN FOUND A WORN BRAKE CASTER THAT CAUSED THE BRAKE NOT TO HOLD. THE BRAKE CASTER WAS REPLACED TO REPAIR THE BED. THE AFFINITY SERVICE MANUAL (MAN013RE) DOCUMENTS, A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER, AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY.

Description of Event or Problem · 1

ACCOUNT ALLEGED THE BRAKES WOULD NOT HOLD ON THIS BED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM RITTER 3600

Patients

Seq Age Sex Outcome Treatment
1 NA