- Manufacturer
- HILL-ROM RITTER
- Product Code
- HDD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WOULD NOT HOLD ON THE BED. THE HILL-ROM TECH FOUND THE BRAKE/STEER CASTERS WERE NOT ADJUSTED PROPERLY. THE CASTERS WERE ADJUSTED TO REPAIR THE BED. THE AFFINITY SVC MANUAL (MAN013RE) DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY.
Description of Event or Problem · 1
ACCOUNT ALLEGED THE BRAKES WOULD NOT HOLD ON THIS BED.