FDA Adverse Event Malfunction Summary report: N

BD ULTRA FINE PEN NEEDLES

MDR report key: 10131488 · Received June 8, 2020

Report

Report Number
9616656-2020-00517
Event Type
Malfunction
Date Received
June 8, 2020
Date of Event
May 28, 2020
Report Date
May 29, 2020
Manufacturer
BECTON DICKINSON AND CO.
Product Code
FMI
UDI-DI
00382903201228
PMA / PMN Number
K162516
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. (B)(4). INVESTIGATION SUMMARY: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 3RD RELATED COMPLAINT FOR NEEDLE CLOG ON LOT # 9232981. A REVIEW OF RISK MANAGEMENT DOCUMENT (B)(4) REVISION 18, INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (PEN NEEDLE, NEEDLE CLOG) WAS CAPTURED AND ADDRESSED APPROPRIATELY. NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSULIN DOES NOT FLOW DURING INJECTION WITH A BD ULTRA FINE¿ PEN NEEDLES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT INSULIN DOES NOT FLOW WHEN INJECTED. STATED, HE DOES NOT PRIME BEFORE INJECTION

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
590869 BD ULTRA FINE PEN NEEDLES HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON AND CO. 320122 9232981 00382903201228

Patients

Seq Age Sex Outcome Treatment
1 Other