FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 1012860
·
Received March 12, 2008
Report
- Report Number
- 3002158293-2008-00114
- Event Type
- Malfunction
- Date Received
- March 12, 2008
- Date of Event
- February 25, 2008
- Report Date
- March 11, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK HAD BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE BATTERY PACK NOT POWERING UP WAS DUE TO A BLOWN FUSE. THE BATTERY PACK WAS REPAIRED, RETESTED AND RESTOCKED. THE ROOT CAUSE OF THE BLOWN FUSE CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY RANDOM COMPONENT FAILURE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED BATTERY PACK. THE LAST PT TO USE THIS BATTERY PACK DID NOT REPORT ANY PROBLEMS.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF BATTERY PACK WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT THE BATTERY PACK WOULD NOT CHARGE. THE LAST PT TO USE THIS BATTERY PACK DID NOT REPORT ANY PROBLEMS WITH IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |