GENERIC YELLOW ISOLATION GOWN WITH CARBON AND TIES AT NECK AND WAIST
Report
- Report Number
- 0001527185-2020-00002
- Event Type
- Malfunction
- Date Received
- June 8, 2020
- Date of Event
- June 1, 2020
- Report Date
- June 30, 2020
- Manufacturer
- STANDARD TEXTILE CO., INC.
- Product Code
- OEA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS NOT AVAILABLE BUT A REPRESENTATIVE SAMPLE OF CIRCULATING INVENTORY WAS PROVIDED. EVALUATION SUMMARY: AN AAMI LEVEL 1 GOWN IS NOT THE RIGHT LEVEL OF PROTECTION WHEN HIGHER VOLUMES OF FLUIDS AND LEANING/PRESSURE IS EXPECTED. THESE CONDITIONS TYPICALLY REQUIRE AAMI LEVEL 2 OR HIGH PERFORMANCE. AAMI LEVEL 1 GOWNS ARE ONLY REQUIRED TO PROVIDE <4.5 GRAMS OF IMPACT RESISTANCE (SIMULATING SPRAY OR SPLATTERING), NO HYDROSTATIC TESTING. INTENDED USE OF AAMI 1 LEVEL PRODUCTS IS WHEN MINIMAL LIQUID CONTACT IS EXPECTED AND NO PRESSING, LEANING OR PRESSURE. CONCLUSIONS: 1. PURCHASED (NEW) PRECAUTION GOWN #6699458P MET SPECIFICATIONS. 2. LAUNDERED PERFORMANCE OF CIRCULATING INVENTORY WAS LOWER THAN EXPECTED BUT STILL MET AAMI LEVEL 1 PERFORMANCE (0.2 G IMPACT PENETRATION). 3. EXPOSURE CONDITIONS INCLUDING HIGHER VOLUMES OF FLUIDS AND LEANING/PRESSURE TYPICALLY REQUIRE AN AAMI LEVEL 2 OR 3 GOWN, I.E., MATERIALS NEED HYDROSTATIC RESISTANCE. LEVEL 1 GOWNS DO NOT HAVE A REQUIREMENT FOR HYDROSTATIC RESISTANCE AND TESTING.
6/3/2020 - INITIATED REQUEST WITH USER FACILITY FOR ADVERSE EVENT DETAILS (PATIENT INFORMATION, 3RD PARTER RE-PROCESSOR AND EXPOSURE CIRCUMSTANCES) AND REQUESTED PRODUCTS FOR EVALUATION/TESTING.
WHILE CLEANING A C-DIFF PATIENT WHO HAD STOOLED THEMSELVES, MOISTURE WICKED THROUGH THE GOWN AT THE LEG AREA WHILE STAFF WAS LEANING/PRESSING AGAINST SOILED BED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 590698 | GENERIC YELLOW ISOLATION GOWN WITH CARBON AND TIES AT NECK AND WAIST | ISOLATION GOWN | OEA | STANDARD TEXTILE CO., INC. | 6699458P | 1156379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |