OLYMPUS
Report
- Report Number
- 8010047-2008-00041
- Event Type
- Malfunction
- Date Received
- March 15, 2008
- Date of Event
- February 13, 2008
- Report Date
- February 14, 2008
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LQC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE OLYMPUS LITHOTRIPTOR UTILIZED IN THE PROCEDURE WAS RETURNED TO OLYMPUS FOR INVESTIGATION. THE INVESTIGATION CONFIRMED THE USER'S REPORT OF THE DAMAGED LITHOTRIPTOR. ONE OF THE FOUR BASKET WIRES WAS OBSERVED TO BE FRACTURED, BUT REMAINED ATTACHED TO THE DISTAL TIP OF THE BASKET. THE CAUSE OF THE USER'S EXPERIENCE WAS DETERMINED TO BE DUE TO THE BILIARY STONE BEING EXCESSIVELY HARD AND LARGE. THIS REPORT IS BEING FILED AS AN MDR IN AN ABUNDANCE OF CAUTION.
A USER FACILITY REPORTED THAT A PT HAD BEEN TRANSPORTED TO THEIR FACILITY WITH ANOTHER CO'S BASKET WIRE LITHOTRIPTOR LODGED AROUND A VERY HARD BILIARY STONE. THE USER FACILITY PERFORMED A THERAPEUTIC ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) TO REMOVE THE STONE AND FREE THE WIRE BASKET. THE NON-OLYMPUS LITHOTRIPTOR DEVICE WAS SUCCESSFULLY REMOVED, AND THE USERS THEN ATTEMPTED TO REMOVE THE STONE USING AN OLYMPUS DEVICE, HOWEVER, THE USERS REPORTED THAT THE STONE WAS REPORTEDLY TOO "CONCRETE" AND A THREAD OF ONE OF THE FOUR BASKET WIRES UNRAVELLED BUT REMAINED ATTACHED TO THE DISTAL TIP OF THE BASKET. THE LITHOTRIPTOR WAS REMOVED FROM THE PT AND ANOTHER DEVICE LITHOTRIPTOR WAS UTILIZED. UPON THIS ATTEMPT, A LOOSE STRAND OF ONE OF THE BASKET WIRES INSIDE THE SHEATH REPORTEDLY FRACTURED, BUT THE USERS WERE ABLE TO CRUSH THE STONE, REMOVE THE LITHOTRIPTOR, AND COMPLETE THE PROCEDURE. THE USER FACILITY REPORTED THAT NO PORTIONS OF THE DEVICES FELL INSIDE THE PT AND THERE WAS NO PT INJURY. IN ADDITION, THE USER FACILITY STATED THAT THE CAUSE OF THIS PHENOMENON WAS DUE TO THE NATURE OF THE BILIARY STONE BEING SO HARD AND LARGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS | SINGLE USE MECHANICAL LITHOTRIPTOR V | LQC | OLYMPUS MEDICAL SYSTEMS CORPORATION | BML-V442QF | 74K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |