FDA Adverse Event Malfunction Summary report: N

STARCLOSE VASCULAR CLOSURE SYSTEM

MDR report key: 1012474 · Received March 12, 2008

Report

Report Number
2953144-2008-00130
Event Type
Malfunction
Date Received
March 12, 2008
Date of Event
January 31, 2008
Report Date
February 20, 2008
Manufacturer
ABBOTT VASCULAR - REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS RETURNED FULLY CLIP DEPLOYED. THE EXTERNAL AND INTERNAL COMPONENTS WERE IN THE CORRECT POST DEPLOYED POSITION. NO STRIKE MARKS WERE FOUND ON THE GARAGE AND PUSHER BLOCKS TO INDICATE INTERFERENCE OF THE SAFETY RELEASE ROD DISTAL END. THE VESSEL BUTTON ENGAGED WITH THE SAFETY RELEASE NUT 10 OF 10 ATTEMPTS. NO DAMAGE, ABNORMALLY OR MFG ISSUE WAS DETECTED. THE DEVICE CONFORMED TO SPECIFICATION. THE ROOT CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCT ANY FINDINGS RELEVANT TO THIS INVESTIGATION. ADDITIONALLY, THE 690 UNUSED, STERILE DEVICES OF THE SAME LOT WERE RETURNED. EVALUATION OF REPRESENTATIVE SAMPLES FOUND NO MALFUNCTION. ALL DEVICES CONFORMED TO SPECIFICATION AND PASSED FUNCTIONAL TESTING WITH ACCEPTABLE RESULTS.

Description of Event or Problem · 1

DEVICE MALFUNCTION: PREMATURE COLLAPSE OF VESSEL LOCATOR WINGS. TIME OF MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: NONE. IT WAS REPORTED THAT A PHYSICIAN, TRAINED IN THE USE OF THE STARCLOSE DEVICE, ATTEMPTED COMMON FEMORAL ARTERIOTOMY CLOSURE AFTER AN UNSPECIFIED PROCEDURE. DURING THUMB ADVANCEMENT, THE VESSEL LOCATOR BUTTON RELEASED RESULTING IN THE VESSEL LOCATOR WINGS PREMATURELY COLLAPSING AND VESSEL ACCESS LOSS. HEMOSTASIS WAS ACHIEVED WITH MANUAL COMPRESSION. THERE WAS NO ADVERSE PT EFFECT. THOUGH REQUESTED, NO ADDITIONAL INFO WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR - REDWOOD CITY NA 52095-6H

Patients

Seq Age Sex Outcome Treatment
1 UNK