FDA Adverse Event Malfunction Summary report: N

SPRINTER RX BALLOON CATHETER

MDR report key: 1012462 · Received March 12, 2008

Report

Report Number
2953200-2008-00125
Event Type
Malfunction
Date Received
March 12, 2008
Date of Event
February 14, 2008
Report Date
February 14, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
LOX
PMA / PMN Number
P970017
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

A 2.0 MM X 15 MM SPRINTER RX DILATATION BALLOON CATHETER WAS USED TO PRE-DILATE AN OM LESION. THE LESION MORPHOLOGY WAS REPORTED TO BE NON-TORTUOUS WITH MODERATE CALCIFICATION AND 95% STENOSIS. THE PHYSICIAN INSERTED THE SPRINTER TO THE LESION SITE. UPON FIRST INFLATION UP TO 4 ATM THE BALLOON WOULD NOT INFLATE. THE SPRINTER BALLOON CATHETER WAS REMOVED SUCCESSFULLY AND POBA COMPLETED THE CASE. NO CLINICAL SEQUELAE WERE REPORTED AND THE PT IS REPORTED TO HAVE EXPIRED DUE TO A NON CARDIAC REASON. THE PHYSICIAN CONFIRMED THE PT WAS SUFFERING FROM MULTIPLE DISORDERS. EVALUATION SUMMARY: MEDTRONIC HAS RECEIVED THE DEVICE AND ITS ANALYSIS HAS BEEN COMPLETED. AS THE BALLOON PROTECTOR AND STYLETTE WERE REMOVED, AND IT WAS EVIDENT THAT THE STYLETTE HAD BEEN PUSHED UP AGAINST THE DISTAL TIP CAUSING DAMAGE TO THE TIP. THE DISTAL TIP WAS BUNCHED AND THE DISTAL END WAS FLARED. NEGATIVE PURGE INDICATED THE PRESENCE OF A BALLOON LEAK. THE BALLOON WAS INFLATED AND TWO LEAKS WERE NOTED ON THE BALLOON. THE LEAKS WERE LOCATED DISTAL TO THE DISTAL MARKER BAND AND DISTAL TO THE PROXIMAL MARKER BAND ON THE OUTSIDE OF THE FOLD. BOTH LEAKS WERE SMALL RADIAL TEARS. PLEASE NOTE THAT THIS DEVICE IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINTER RX BALLOON CATHETER 74LOX LOX MEDTRONIC CARDIOVASCULAR GALWAY NA 0000606262

Patients

Seq Age Sex Outcome Treatment
1 UNK