FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-715NAS PRDGM INS V2.1 SK EN
MDR report key: 1011907
·
Received March 10, 2008
Report
- Report Number
- 2032227-2008-00463
- Event Type
- Malfunction
- Date Received
- March 10, 2008
- Date of Event
- February 25, 2008
- Report Date
- February 25, 2008
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TESTS, DUE TO AN OUT OF PHASE ENCODER SIGNAL. NO MOTOR ERROR ALARMS WERE NOTED.
Description of Event or Problem · 1
THE CUSTOMER CALLED TO REPORT MOTOR ERROR ALARMS ON THE INSULIN PUMP. TROUBLESHOOTING WAS PERFORMED AND THE INSULIN PUMP WAS STUCK IN THE REWIND MODE. THE CUSTOMER WAS UNABLE TO CLEAR THE INSULIN PUMP FROM THAT MODE. ADVISED THE CUSTOMER TO DISCONTINUE USING THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-715NAS PRDGM INS V2.1 SK EN | INFUSION PUMP | LZG | MEDTRONIC MINIMED | MMT-715NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |