FDA Adverse Event Injury Summary report: N

COZMO INSULIN PUMP

MDR report key: 1011876 · Received March 10, 2008

Report

Report Number
2183502-2008-00035
Event Type
Injury
Date Received
March 10, 2008
Date of Event
February 11, 2008
Report Date
March 6, 2008
Manufacturer
SMITHS MEDICAL MD (FORMERLY DELTEC, INC.)
Product Code
LZG
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. DELIVERY AND ACCURACY TESTS WERE PERFORMED, THE DEVICE WAS FOUND TO PASS ALL DELIVERY AND ACCURACY TESTS. THE PUMP HISTORY WAS DOWNLOADED AND ANALYZED NO ANOMALIES WERE FOUND. NO OPERATIONAL FAILURE WAS DETECTED. THE LCD SCREEN WAS FOUND TO BE CRACKED AND DAMAGED.

Description of Event or Problem · 1

INFO WAS RECEIVED THAT REPORTED A PT WAS HOSPITALIZED IN 2008 DUE TO AN INCIDENT OF DIABETIC KETOACIDOSIS. IT WAS REPORTED THAT THE PT WOKE UP FEELING ILL AND HE HAD BLOOD GLUCOSE OF >450MG/DL. THE PT WAS BROUGHT TO THE HOSPITAL AND HAD BLOOD GLUCOSE OF 511MG/DL UPON ADMISSION. THE PT WAS TREATED WITH IV FLUIDS AND INSULIN. THE DEVICE SHOULD BE RETURNED FOR EVALUATION; TO ENSURE THAT IT IS FUNCTIONING CORRECTLY AND DID NOT CONTRIBUTE TO THE REPORTED INCIDENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COZMO INSULIN PUMP INSULIN PUMP LZG SMITHS MEDICAL MD (FORMERLY DELTEC, INC.) 1700 NA

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization